Device Attributes Description

Lakeshore_331 Class

Revision: release_1_0_2 - Author: coquet

Scalar Attributes
Attribute name
Data Type
R/W Type
Expert
temperature: temperature voie principale en Kelvin read : temperature courante write : nouvelle consigne Attribut polle pour controler si la consigne est atteinte polling par defaut : 250 ms (si on passe 3 fois par consigne = temperature sans depassement d'1 Delta Temperature)
DEV_DOUBLE
READ_WRITE
No
output: output value of regulation from 0 to 100 %
DEV_DOUBLE
READ
Yes
range: current range can be Low(0.01) Medium(0.1) High(1)
DEV_STRING
READ
Yes
deadBand: dead band in witch we consider that setpoint is reached : if dead band is 0.5 e.g. if setpoint is 12 �K and temp is 12.4 -> OK if setpoint is 12 �K and temp is 11.6 -> OK if setpoint is 12 �K and temp is 11.4 -> KO
DEV_DOUBLE
READ_WRITE
Yes
inputA: temperature in Kelvin for Input A
DEV_DOUBLE
READ
Yes
inputB: temperature in Kelvin for Input B
DEV_DOUBLE
READ
Yes
loop1Input: input selected for loop 1 by the commands : Loop1SelectInputA Loop1SelectInputB
DEV_STRING
READ
No