TANGO
Device Server



Monochromator for the Tempo Beam Line
Device Attributes Description

MonochromatorTempo Class

Revision: release_1_1_2 - Author: postollec

Scalar Attributes
Attribute name
Data Type
R/W Type
Expert
currentSlotName: The current slot name
DEV_STRING
READ
No
currentBehaviour: The current choosen behaviour.
DEV_STRING
READ
No
energy: The value of the energy of the monochromator
DEV_DOUBLE
READ_WRITE
No
energyRenishaw: The value of the energy encoder of the monochromator (from the encoder whic can be only read)
DEV_DOUBLE
READ
No
lambda: The value of the wavelenght lambda.
DEV_DOUBLE
READ_WRITE
No
alpha: The alpha angle which represents the angle between the beam line and the network.
DEV_DOUBLE
READ_WRITE
No
beta: The beta angle which represents the angle between the network and the beam from the output of the network.
DEV_DOUBLE
READ
No
mirrorDeviationAngle: The deviation angle of the mirror.
DEV_DOUBLE
READ
No
currentResolutionCriterion: The current resolution criterion name (Fixed Resolution or Fixed Slits). If the Fixed Resolution criterion is choosen, the value of the attribute called resolution is used for the computing.
DEV_STRING
READ
No
currentVGDCriterion: The current VGD criterion name
DEV_STRING
READ
No
currentConstraint: The current constraint value. In fact this is the last value used during during an energy write.
DEV_DOUBLE
READ
No
encoderConstraint: The value of the constraint according the position of the motors.
DEV_DOUBLE
READ
No
resolution: The current resolution
DEV_DOUBLE
READ
No
resolvingPower: The resolving power value (E/DeltaE)
DEV_DOUBLE
READ
No
selectedResolution: The selected resolution value. If we write a new value, the current resolution criterion is set to Fixed Resolution
DEV_DOUBLE
READ_WRITE
No
selectedResolvingPower: The selected resolving power value. If we write a new value, the current resolution criterion is set to Resolving Power Criterion
DEV_DOUBLE
READ_WRITE
No
currentGratingSize: The current grating value in lines by millimeters.
DEV_DOUBLE
READ
No
currentGratingName: The name of the current grating of the current slot
DEV_STRING
READ
No
relativeGratingPosition: The relative position over the grating.
DEV_DOUBLE
READ_WRITE
No
currentGrooveDepth: The current groove depth of the grating
DEV_DOUBLE
READ
No
optimalHighFluxX: Give the optimal X VGD position for the current Energy for the High Flux criterion
DEV_DOUBLE
READ
No
optimalHighRejectionX: Give the optimal X VGD position for the current Energy for the High Harmonics Rejection
DEV_DOUBLE
READ
No
currentMirrorName: The current mirror name of the current slot.
DEV_STRING
READ
No
entrySlitImage: The image of the entry slit over the exit slit space.
DEV_DOUBLE
READ
No
gratingPosition: The grating position value (for the VGD and the choice of the grating)
DEV_DOUBLE
READ_WRITE
Yes
mirrorPosition: The mirror position
DEV_DOUBLE
READ_WRITE
Yes
exitSlitAperture: The exit slit aperture value
DEV_DOUBLE
READ_WRITE
Yes
entrySlitAperture: The entry Slit Aperture
DEV_DOUBLE
READ_WRITE
Yes
selectedExitSlitAperture: The selected exit slit aperture used to change to fixed slit criterion. This value is applied to the exit slit aperture before using the fixed slits criterion.
DEV_DOUBLE
READ_WRITE
No







ESRF - Software Engineering Group