TurboPumpAL
Device States Description

TurboPumpAL Class

Revision: release_1_0_0 - Author: le



States
Names
Descriptions
ON Turbo pump is ON If at least one of the following bits is TRUE: BSV_Power BSV_Start BSV_Acceleration BSV_Nominal BSV_Braking BSV_AtSpeed
OFF Turbo pump is OFF If BSV_Start is FALSE
FAULT The pump is in FAULT if DSV_EtatAlarmes <>0 or BSV_Fault is TRUE Note: STATUS should show the folowwing text associated with 19 Alarms: Bit number (not numerical value!) / Alarm type 0 (LSB) / Fault: Seized pump 1 / Alert: Power overheating 2 / Alert: Motor overheating 3 / Alert: Motor control 4 / Fault: Hall sensor 5 / Fault: Overcurrent 6 / Fault: External safety 7 / Fault: Magnetic suspension 8 / Alert: Power voltage 9 / Alert: Power current 10 / Fault: Bearing to be changed 11 / Alert: ACT temperature 12 / Alert: Pump temperature 13 / Fault: Sensor 1 temperature 14 / Fault: No connect 15 / Fault: EEPROM check 16 / Fault: Input power 17 / Fault: No coding 18 / Fault: Not connected 19 / Fault: Sensor 2 temperature Others bits not significant
UNKNOWN Communication is broken
STANDBY Read from BSV_Standby
ALARM