States |
Names |
Descriptions |
ON |
Turbo pump is ON
If at least one of the following bits is TRUE:
BSV_Power
BSV_Start
BSV_Acceleration
BSV_Nominal
BSV_Braking
BSV_AtSpeed |
OFF |
Turbo pump is OFF
If BSV_Start is FALSE |
FAULT |
The pump is in FAULT if DSV_EtatAlarmes <>0 or BSV_Fault is TRUE
Note: STATUS should show the folowwing text associated with 19 Alarms:
Bit number (not numerical value!) / Alarm type
0 (LSB) / Fault: Seized pump
1 / Alert: Power overheating
2 / Alert: Motor overheating
3 / Alert: Motor control
4 / Fault: Hall sensor
5 / Fault: Overcurrent
6 / Fault: External safety
7 / Fault: Magnetic suspension
8 / Alert: Power voltage
9 / Alert: Power current
10 / Fault: Bearing to be changed
11 / Alert: ACT temperature
12 / Alert: Pump temperature
13 / Fault: Sensor 1 temperature
14 / Fault: No connect
15 / Fault: EEPROM check
16 / Fault: Input power
17 / Fault: No coding
18 / Fault: Not connected
19 / Fault: Sensor 2 temperature
Others bits not significant |
UNKNOWN |
Communication is broken |
STANDBY |
Read from BSV_Standby |
ALARM |
|